Note: • If the value measured by the instrument is not

close to the selected buffer, “WRONG” “

2

BUF

”

and “WRONG” “ ” tags will blink alternately.

In this case, check if the correct buffer has been

used or regenerate the electrode by following the Cleaning Proce-

dure (see page 45). If necessary, change the buffer or the

electrode.

• If “WRONG”, “Buffer pH” tags and “Old” tag on secondary LCD are

displayed blinking, an inconsistency between new and previous

(old) calibration is detected. Clear calibration parameters by

pressing LOCK/CLR and proceed with calibration from the

current calibration point (the instrument will keep all confirmed

values during current calibration).

• If LOCK/CLR is pressed at the first calibration point, the instru-

ment clears all calibration parameters and returns to measurement

mode.

• The “WRONG” tag and temperature value are displayed blinking

if temperature reading is out of the defined temperature range of

the buffer. Calibration cannot be confirmed in this situation.

• Press RANGE to display the temperature reading

on the LCD during calibration.

TWO-POINT CALIBRATION
• Proceed as described in “Three-point calibration” section.

• Press CAL after the second calibration point was confirmed.

The instrument will return to measurement mode and will memorize the

two-point calibration data.
ONE-POINT CALIBRATION
• Proceed as described in “Three-point calibration” section.

• Press CAL after the first calibration point was confirmed.

The instrument will return to measurement mode and will memorize the

one-point calibration data.

Note: Calibration parameters are evaluated taking in consideration the new

values for the confirmed buffers in current calibration and the old

values, if existing, for the other buffers. To clear old calibration values,

press LOCK/CLR before exiting calibration.