Hanna Instruments HI 255 User Manual

Page 15

background image

15

RELATIVE mV CALIBRATION

• Press CAL when the instrument is in RELATIVE mV measurement mode.

The “mV” and “ ” tags will blink. Absolute mV is displayed on the

primary LCD and “AbS” message is displayed on the secondary LCD.

• When the reading is stable and inside the offset window (±2000 mV), the

instrument asks for confirmation.

• If the reading is out of the offset window, “WRONG” tag will be displayed.

• Press CFM to confirm absolute value. The instrument will display 0.0 mV

on the primary LCD and “rEL” message on the secondary LCD line. In this

moment the relative mV offset is equal to absolute mV reading.

• Use the ARROW keys if you want to change the displayed value.

• Press CFM to confirm relative mV calibration. The instrument will return to

measurement mode.