• EC calibration data (if available), which contains:

• the number of calibrated standards (1 char)
• the offset factor, with sign and decimal point

(7 chars)

• the cell constant, with sign and decimal point

(7 chars)

• the calibration time, yymmddhhmmss (12 chars)
• standards information (for each standard)

• standard value, with sign and decimal point
(7 chars).
• buffer unit (2 chars; 00-μS; 01-mS)
• Reference Temperature with and decimal point

(4 chars)

• Temperature Compensation mode (2 chars)

• 00 - no temperature compensation
• 01 - automatic temperature compensation
• 00 - manual temperature compensation

• TC coeficient with sign and decimal point (4 chars)
• calibration time, yymmddhhmmss (12 chars).

• Na Cl Calibration data

• the number of calibrated data (1 char)
• salinity coeficient, with sign and decimal point

(7 chars)

• Cell constant, with sign and decimal point

(2 chars)

• the calibration time (2 chars)
• buffer information, for each buffer:
• Reference Temperature with and decimal point

(4 chars)

• Temperature Compensation mode (2 chars)

• 00 - no temperature compensation
• 01 - automatic temperature compensation
• 00 - manual temperature compensation

• TC coeficient with sign and decimal point (4 chars)
• calibration time, yymmddhhmmss (12 chars).

PAR Requests the setup parameters setting.
The answer string contains:

• Instrument ID (4 chars)