Hanna Instruments HI 2550 User Manual
Page 10

10
• If LOCK key was pressed and the reading goes out of range, the
full-scale value of the frozen range will be displayed blinking.
The conductivity reading is affected by temperature.
Three options for temperature compensation are available in conductivity
measurement mode.
Automatic (ATC): The conductivity probe has a built-in temperature
sensor; the temperature value is used to automatically compensate the
EC/TDS reading.
Manual (MTC): The temperature value, shown on the secondary LCD,
can be manually set with the ARROW keys. The “ºC” tag blinks when
this option is active. This value will be used to compensate the EC/TDS
reading.
No Compensation (notc): The temperature value is displayed, but not
taken into account. When this option is selected the “ºC” tag will blink
with slower frequency. The reading displayed on the primary LCD is the
uncompensated EC or TDS value.
To select the desired option, press the
ALT&ATC keys until the option is displayed
on the LCD.
Notes:• The default compensation mode is ATC.
•
If no temperature probe is detected, ATC mode can not be
selected and the instrument displays “----” on the secondary LCD.
If temperature compensation is selected, measurements are compensated
using the temperature coefficient (default value 1.90 %/ºC). To change
the temperature coefficient, enter the setup mode and select “tc” (see
SETUP for details, page 36). The current temperature coefficient can be
quickly viewed by pressing the ALT&TC keys. The value is briefly
displayed on the secondary LCD.
• If the temperature reading exceeds the limits of the meter (–20.0 ºC
to 120.0 ºC), the “ºC” tag will blink and the closest full scale value
will be displayed.