Cyanuric Acid
CYANURIC ACID
SPECIFICATIONS
Range
0 to 200 mg/L
Resolution
1 mg/L from 0 to 100 mg/L;
10 mg/L above 100 mg/L
Accuracy
±1 mg/L ±15% of reading at 25 °C
Typical EMC
±1 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Adaptation of the turbidimetric method. The reaction between cyanuric acid and the
reagent causes a white suspension in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93722-0
Powder reagent
1 packet
REAGENT SETS
HI 93722-01 Reagents for 100 tests
HI 93722-03 Reagents for 300 tests
For other accessories see page 34.
MEASUREMENT PROCEDURE
• Select the
Cyanuric Acid method using the procedure described in the Method Selection section (see page 15).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for
measurement.
10 mL
- pH 213 (17 pages)
- pH 211R (13 pages)
- pH 21 (9 pages)
- pH 209V (9 pages)
- pH 209R (13 pages)
- HI 99181 (2 pages)
- HI 99171 (2 pages)
- HI 99163N (2 pages)
- HI 99161N (2 pages)
- HI 99141 (2 pages)
- HI 99131 (2 pages)
- HI 99121 (9 pages)
- HI 99104 (2 pages)
- HI 991003 (2 pages)
- HI 991002 (2 pages)
- HI 991001 (2 pages)
- HI 98402 (9 pages)
- HI 98401 (8 pages)
- HI 98308 (2 pages)
- HI 98240 (21 pages)
- HI 98203 (2 pages)
- HI 98202 (2 pages)
- HI 98201 (2 pages)
- HI 98185 (72 pages)
- HI 98172 (52 pages)
- HI 98160N (48 pages)
- HI 98160 (48 pages)
- HI 98150 (17 pages)
- HI 9815 (2 pages)
- HI 981410 (2 pages)
- HI 981408 (2 pages)
- HI 98121 (2 pages)
- HI 98120 (2 pages)
- HI 98113 (2 pages)
- HI 98110 (2 pages)
- HI 98108 (2 pages)
- HI 98107 (2 pages)
- HI 98106 (2 pages)
- HI 98103 (2 pages)
- HI 96108 (2 pages)
- HI 96107 (2 pages)
- HI 96106 (2 pages)
- HI 96104 (28 pages)
- HI 96101 (36 pages)
- HI 9321 (36 pages)