SPECIFICATIONS

Range

0 to 500 mg/L (as CaCO

3

)

Resolution

5 mg/L

Accuracy

±5 mg/L ±10% of reading at 25 °C

Typical EMC

±5 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Colorimetric Method

. At different alkalinity levels a distinctive range of colors from yellow

to green and greenish blue will develop.

REQUIRED REAGENTS

Code

Description

Quantity/test

HI 93755-0

Alkalinity Indicator Reagent

1 packet

REAGENT SETS

HI 93755-01 Reagents for 100 tests
HI 93755-03 Reagents for 300 tests
For other accessories see page 34.

MEASUREMENT PROCEDURE

• Select the

Alkalinity method using the procedure described

in the

Method Selection section (see page 15).

• Fill the cuvette with 10 mL of unreacted sample, up to the

mark, and replace the cap.

• Place the cuvette into the holder and close the lid

• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

ALKALINITY

Alkalinity

10 mL