SPECIFICATIONS

Range

0.00 to 5.00 mg/L

Resolution

0.01 mg/L from 0.00 to 2.50 mg/L;
0.10 mg/L above 2.50 mg/L

Accuracy

±0.03 mg/L ±3% of reading at 25 °C

Typical EMC

±0.01 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the USEPA method 330.5 and

Standard Methods for the Examination

of Water and Wastewater, 20

th

edition, 4500-Cl G. The reaction between free chlorine

and the DPD reagent causes a pink tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93701-0

DPD powder Reagent

1 packet

REAGENT SETS

HI 93701-01 Reagents for 100 tests
HI 93701-03 Reagents for 300 tests
For other accessories see page 34.

MEASUREMENT PROCEDURE

• Select the

Free Chlorine method using the procedure described in the Method Selection section (see

page 15).

• Fill the cuvette with 10 mL of unreacted sample (up to the

mark) and replace the cap.

• Place the cuvette into the holder and close the lid.

Free Chlorine

FREE CHLORINE

10 mL