Total Chlorine

TOTAL CHLORINE

SPECIFICATIONS

Range

0.00 to 5.00 mg/L

Resolution

0.01 mg/L from 0.00 to 2.50 mg/L;
0.10 mg/L above 2.50 mg/L

Accuracy ±0.03 mg/L ±3% of reading at 25 °C
Typical EMC

±0.01 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the EPA method 330.5 and

Standard Methods for the Examination of

Water and Wastewater, 20

th

edition, 4500-Cl G. The reaction between chlorine and the

DPD reagent causes a pink tint in the sample.

REQUIRED REAGENTS

POWDER:

Code

Description

Quantity

HI 93711-0

DPD Powder Reagent 1 packet

REAGENT SETS

HI 93711-01 Reagents for 100 tests
HI 93711-03 Reagents for 300 tests
For other accessories see page 34.

MEASUREMENT PROCEDURE

• Select the

Total Chlorine method using the procedure described in the Method Selection section (see page 15).

• Fill the cuvette with 10 mL of unreacted sample (up to the mark)

and replace the cap.

• Place the cuvette into the holder and close the lid.