beautypg.com

Hanna Instruments HI 96711 User Manual

Page 9

background image

16

17

• After a few seconds, the display will

show “-0.0-”. The meter is now zeroed
and ready for measurement.

• Remove the cuvette.

• Add the content of one packet of the

specific test reagent, for:
Free Chlorine:

1 packet of HI 93701-0

or

Total Chlorine:

1 packet of HI 93711-0

• Replace the cap and shake gently for 20

seconds (or 2 minutes in case of seawater
analysis).

• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/ /TIMER

for three

seconds.

The display will show the

countdown prior to measurement. An
audible “beep” indicates the end of
countdown period.

• Alternatively, wait for 1 minute for free

chlorine or 2’30’’ for total chlorine and
just press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

To compensate the meter for the sample turbidity or color, the measurement takes place in two
phases. First, the meter is zeroed using the unreacted sample. After the reagents are added the
reacted sample is measured.
Important note:

Free and Total Chlorine have to be measured separately with fresh unreacted

samples if both values are requested.

MEASUREMENT PROCEDURE

• Turn the meter on by pressing ON/OFF.

The display briefly shows all tags on.

• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.
The blinking “ZERO” indicates that the
instrument needs to be zeroed first.

• Pay attention to the selected range.

To change the range, simply press
RANGE/GLP/

.

• Fill the cuvette with 10 mL of unreacted

sample, up to the mark, and replace the
cap.

• Place the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.