To display the calibration date:

• Press and hold for three seconds

RANGE/GLP/

to enter

GLP mode. The

calibration month and day will appear on
the main display and the year on the
secondary display.

• If no calibration was performed, the

factory calibration message, “F.CAL” will
appear on the main display and the
instrument returns to

measurement mode

after three seconds.

It is possible to delete the calibration and
restore factory calibration.

• Press and hold for three seconds

RANGE/GLP/

to enter

GLP mode.

• Press READ/ /TIMER to enter in the

factory calibration restore screen. The
instrument asks for confirmation of user
calibration delete.

• Press ZERO/CFM to restore the factory

calibration or press RANGE/GLP/ again
to abort factory calibration restore.

• The instrument briefly indicates “donE”

upon restoration of factory calibration prior
to returning to

measurement mode.

FACTORY CALIBRATION RESTORE

In

GLP mode, the last user calibration date can be verified and the factory calibration can be

restored.

LAST CALIBRATION DATE

G L P

• Press RANGE/GLP/ to edit the desired

year (2009-2099). If the key is kept
pressed, the year number is automatically
increased.

• When the correct year has been set, press

ZERO/CFM

or READ/ /TIMER to

confirm. Now the display will show the
month blinking.

• Press RANGE/GLP/ to edit the desired

month (01-12). If the key is kept pressed
the month number is automatically
increased.

• When the correct month has been set,

press ZERO/CFM or READ/ /TIMER
to confirm. Now the display will show
the day blinking.

• Press RANGE/GLP/ to edit the desired

day (01-31). If the key is kept pressed
the day number is automatically increased.

Note:

It is possible to change the editing
from day to year and to month by
pressing READ/ /TIMER.

• Press ZERO/CFM to save the calibration

date.

• The instrument displays “Stor” for one

second and the calibration is saved.

• The instrument will return automatically

to

measurement mode by displaying

dashes on the LCD.

DATE INPUT

or

or