Errors and warnings – Hanna Instruments HI 96711 User Manual

Page 6

background image

10

11

ERRORS AND WARNINGS

The instrument shows clear messages when erroneous condition appears. Messages are also displayed
when the obtained values are outside expected range. The beeper is playing a beep on errors.

a) on zero reading

Light High:

There is too much light to perform a measurement.

Please check the preparation of the zero cuvette.

Light Low:

There is not enough light to perform a measurement.

Please check the preparation of the zero cuvette.

No Light:

The instrument cannot adjust the light level. Please

check that the sample does not contain any debris.

Inverted cuvettes:

The sample and the zero cuvette are inverted.

Zero:

A zero reading was not taken. Follow the instructions of

the measurement procedure for zeroing the meter.

b) on sample reading

DISPLAY ELEMENTS DESCRIPTION

1) The measuring scheme (lamp, cuvette, detector), appears during different phases of zero or

reading measurement

2) Error messages and warnings
3) The battery icon indicates the charge state of the battery
4) The hourglass appears when an internal check is in progress
5) Status messages
6) The chronometer appears when the reaction timer is running
7) The month, day and date icons appear when a date is displayed
8) Four digit main display
9) Measuring units

10) Four digit secondary display

10

8

5

2

9

7

3

4

6

1