beautypg.com

Hanna Instruments HI 96711 User Manual

Page 11

background image

20

21

• Remove the cuvette.

• Place the specific CAL CHECK

TM

Standard

Cuvette B into the cuvette holder, for:
Free Chlorine:

B, HI 96701-11

or

Total Chlorine:

B, HI 96711-11

• Ensure that the notch on the cap is

positioned securely into the groove.

• Press READ/ /TIMER and the lamp,

cuvette and detector icons will appear on
the display, depending on the
measurement phase.

• After measurement the instrument will

show for three seconds the CAL CHECK

TM

Standard value.

Note:

If the display shows “STD HIGH”, the
standard value was too high. If the
display shows “STD LOW”, the
standard value was too low. Verify
that both CAL CHECK

TM

Standard

Cuvettes, A and B are free from
fingerprints or dirt and that they are
inserted correctly.

• Then the date of the last calibration (e.g.:

”01.08.2009”

) appears on the display,

or “01.01.2009 if the factory calibration
was selected before. In both cases the
year number is blinking, ready for date
input.

CALIBRATION PROCEDURE

Note:

It is possible to interrupt calibration procedure at any time by pressing CAL CHECK or

ON/OFF

keys.

Warning

: Do not calibrate the instrument with standard solutions other than the HANNA CAL

CHECK

TM

Standards, otherwise erroneous results will be obtained.

When calibrating, only the selected range is affected.

• Turn the meter on by pressing ON/OFF.

• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.

• To change the range, simply press

RANGE/GLP/ .

• Press and hold CAL CHECK

for three

seconds to

enter

calibration mode. The

display will show “CAL” during
calibration procedure. The blinking
“ZERO”

asks for instrument zeroing.

• Place the CAL CHECK

TM

Standard Cuvette

A into the cuvette holder and ensure that
the notch on the cap is positioned securely
into the groove.

• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• After a few seconds the display will show

“-0.0-”

. The meter is now zeroed and

ready for calibration. The blinking
“READ”

asks for reading calibration

standard.