CLEANING PROCEDURE
• General

Soak in HI 7061 general cleaning solution for

approximately 1 hour.

Removal of films, dirt or deposits on the membrane/junction:
• Protein

Soak in Hanna HI 7073 protein cleaning solution

for 15 minutes.

• Inorganic

Soak in Hanna HI 7074 inorganic cleaning solution

for 15 minutes.

• Oil/grease

Rinse with Hanna HI 7077 oil & fat cleaning solution.

IMPORTANT: After performing any of the cleaning procedures, rinse
the electrode thoroughly with distilled water, drain and refill the
reference chamber with fresh electrolyte, (not necessary for gel-filled
electrodes) and soak the electrode in HI 70300 storage solution for at

least 1 hour before taking measurements.

TROUBLESHOOTING
Evaluate your electrode performance based on the following:

••

•• Noise (readings fluctuate up and down) could be due to:

----- Clogged/Dirty Junction: refer to the above "Cleaning Procedure"
----- Loss of shielding due to low electrolyte level (in refillable

electrodes only): refill with HI 7071 solution for single junction

or HI 7082 for double junction electrodes

••

•• Dry Membrane/Junction: soak in HI 70300 storage solution for

at least 1 hour

••

•• Drifting: soak the electrode tip in warm HI 7082 solution for one

hour and rinse the tip with distilled water; refill with fresh electrolyte
(HI 7071 for single junction and HI 7082 for double junction

electrodes)

••

•• Low Slope: refer to the above "Cleaning Procedure"

••

•• No Slope: check the electrode for cracks in glass stem or bulb

(replace the electrode if cracks are found)

••

•• Slow Response/Excessive Drift: soak the tip in HI 7061 solution

for 30 minutes, rinse thoroughly in distilled water and then follow
the above "Cleaning Procedure"

••

•• For ORP Electrodes: polish the metal tip with a light abrasive

paper (pay attention not to scratch the surface) and rinse
thoroughly with water.