Specifications – Hanna Instruments HI847492 User Manual
Page 7

7
Range
0.00 to 9.99 FTU
10.0 to 99.9 FTU
100 to 1000 FTU
Range Selection
Automatically
Resolution
0.01 FTU from 0.00 to 9.99 FTU
0.1 FTU from 10.0 to 99.9 FTU
1 FTU from 100 to 1000 FTU
Accuracy
±2% of reading plus 0.05 FTU
Repeatability
±1% of reading or 0.02 FTU, whichever is greater
Stray Light
< 0.1 FTU
Light Source
LED @ 580nm
Light Detector
Silicon Photocell
Method
Ratio Nephelometric Method.
Display
60 x 90mm backlit LCD
Calibration
Two, three or four point calibration
LOG Memory
200 records
Serial Interface
USB 1.1
Environment
0 to 50°C (32 to 122°F); max 95% RH non-condensing
Power supply
4 x 1.5V AA alkaline batteries or AC adapter
Auto Shut-off
After 15 minutes of non-use
Dimensions
224 x 87 x 77 mm (8.8 x 3.4 x 3.0”)
Weight
512 g (18 oz.)
SPECIFICATIONS