Alternatively, press LOG/CFM to skip the first calibration point.

• Then the LCD will show the second calibration

point (15.0 FTU) and “CAL P.2”, while “READ”
is blinking.
Note: If you’re using different calibration
standards, change the displayed value by
pressing UP or DOWN keys until the display
shows the desired value.

• Remove the <0.1 FTU standard cuvet and

place the 15.0 FTU standard cuvet into the
holder.

• Close the lid and press READ . “----” and the

“LED and Read Status” indicator will blink
again while making the reading.

• At the end of the reading, the third calibration

point (100 FTU) and “CAL P.3” is displayed. If
desired, the value can be changed using the
UP or DOWN keys.
Note: At this moment it is possible to exit
calibration by pressing CAL/RCL. The instrument
will memorize the two-point (<0.1 and 15.0
FTU) calibration data and will return to
measurement mode.

• Remove the 15.0 FTU standard cuvet from the

meter and place the 100 FTU standard cuvet
into the holder.

• Close the lid and press READ . “----” and the

“LED and Read Status” indicator will blink
again while making the reading.

<0.1
FTU

15.0
FTU

15.0
FTU

100

FTU