Hanna Instruments HI 121 User Manual
Page 46

46
SET
Is equivalent to pressing SET
PRT
Is equivalent to pressing PRINT (HI 123 only)
CLR
Is equivalent to pressing CLR
OFF
Is equivalent to pressing OFF
AHD
Is equivalent to pressing AutoHOLD
CHR xx
Change the instrument range according with the parameter
value (xx):
• xx=00 pH range/0.001 resolution, pH channel
• xx=01 pH range/0.01 resolution, pH channel
• xx=03 mV range, pH channel
• xx=04 Relative mV range, pH channel
• xx=13 mV range, ISE channel
• xx=14 Relative mV range, ISE channel
• xx=15 ISE range, ISE channel
The meter will answer for these commands with:
where:
COMMANDS REQUIRING AN ANSWER
RAS
Causes the instrument to send a complete set of readings
in according with the current range:
• pH, mV and temperature reading on pH range.
• mV and temperature reading on mV range.
• Rel mV, absolute mV and temperature reading on
Rel mV range.
Note: The temperature is sent with two decimals and in Celsius degrees,
even if the set temperature unit was Fahrenheit degrees.
DA?
Requests the date.
TI?
Requests the time.
MDR
Requests the instrument model name and firmware code.
GLP
Requests the calibration data record.
PAR
Requests the setup parameters setting.
NSL
Requests the number of logged samples.
?ML
Requests the information about AutoLOG.