• Press

to change the desired method parameters as indicated in the first step of the method.

• Press

to take the second mV reading.

• When the reading is stable, press

to store the second mV reading. The ISE measurement results

will be displayed on the LCD.

• Press

to log the current results into a ISE Method Report or press

to return to

ISE

Measure

mode.

• If pressing

, the method parameters can be changed and the sample concentration will be

recalculated.

• Press

to start another measurement.

Note: Press

at any time to stop the measurement and return to

ISE Measure

mode.

ANALYTE ADDITION
To measure the concentration of a sample using the Analyte Addition incremental method:
• Press

and then

to select

ISE Measure

mode for the selected channel.

• Select the Analyte Addition method (see ISE Setup for details, page 33).
• Press

to start the measurement. The first step of the method will be displayed on the LCD in

which the user is notified to add the appropriate ISA Buffer Volume (if different from 0 mL) to the standard
and to immerse the electrode into the Standard Volume of set concentration. The method parameters are
also displayed on the LCD.

• Press

to set the method parameters. Press

/

to select next/previous parameter

to be edit, then press

and use

or

to set the desired parameter value. Press

to save the modified value and then press

to exit method parameters edit menu.

• Make sure that the set parameters values are the used ones and submerge the Ion Selective Electrode tip

approximately 4 cm (1½”) into the standard solution.

• Press

to take the first mV reading.

• When the reading is stable, press

to store the first mV reading. The second step of the method

will be displayed on the LCD in which the user is notified to add the Sample Volume to the standard
solution. The method parameters are also displayed on the LCD.

• Press

to change the desired method parameters as indicated in the first step of the method.

• Press

to take the second mV reading.

• When the reading is stable, press

to store the second mV reading. The ISE measurement results

will be displayed on the LCD.

• Press

to log the current results into a ISE Method Report or press

to return to

ISE Measure

mode.