pH CALIBRATION & MEASUREMENTS

pH CALIBRATION & MEASUREMENTS

pH CALIBRATION & MEASUREMENTS

pH CALIBRATION & MEASUREMENTS

pH CALIBRATION & MEASUREMENTS

pppppH CALIBRATION

H CALIBRATION

H CALIBRATION

H CALIBRATION

H CALIBRATION

Calibrate the instrument often, especially if high accuracy is required.
The instrument should be recalibrated:
• Whenever the pH electrode is replaced.
• At least once a week.
• After testing aggressive chemicals.
• When “No pH Calibration” or “pH Calibration Expired” message appears on the LCD, in the Reminder messages area.
PREPARATION
Pour small quantities of the buffer solutions into clean beakers. If possible, use plastic beakers to minimize

any EMC interferences.
For accurate calibration and to minimize cross-contamination, use two beakers for each buffer solution. One

for rinsing the electrode and one for calibration.
If you are measuring in the acidic range, use pH 7.01 or 6.86 as first buffer and pH 4.01/3.00 or 1.68 as

second buffer. If you are measuring in the alkaline range, use pH 7.01 or 6.86 as first buffer and pH 10.01/

9.18 or 12.45 as second buffer.
For extended range measurements (acidic and alkaline), perform a five-point calibration by selecting five of

the available buffers.
CALIBRATION PROCEDURE
Calibration has a choice of 8 memorized buffers: pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45 and

up to 5 custom buffers.
For accurate measurements it is recommended to perform a five-point calibration. However, at least a two-point

calibration is suggested.
pH Calibration screen description