Hanna Instruments HI 4221 User Manual

Page 22

background image

22

1st Cal. Point

Two options are available for the 1st Cal. Point parameter: Point and Offset.
If Point option is selected, the slope values adjacent to the calibration points will be reevaluated (normal calibration).
If at least a two-point calibration has been performed and an offset correction of the electrode is wanted
(maintaining the existing slope values), perform a one-point calibration using the Offset option.
To set the 1st Cal. Point:
• Press

while in

pH Measure

mode.

• Press

.

• Use

or

to select the Calibration option.

• Press

and use

or

to highlight the 1st

Cal. Point option.

• Press

and use

or

to highlight the

desired option.

• Press

to confirm your selection or press

to

cancel operation.

Edit Custom Buffers

If you want to use other buffers than the ones already memorized, the Edit Custom Buffers option is available,

allowing you to set the desired pH buffers. Up to five pH custom buffers can be set.
To edit/set the Custom Buffers:
• Press

while in

pH Measure

mode.

• Press

.

• Use

or

to select the Calibration option.

• Press

and use

or

to highlight the Edit

Custom Buffers option.

• For a previous set value, press

to set the custom buffer value

to “----” if desired and confirm the setting by pressing

,

otherwise press

to edit the selected custom buffer.

• While in edit custom buffer menu press

to set the custom

buffer value to 7.000 pH and then use

or

to set

the desired custom buffer value.

This manual is related to the following products: