beautypg.com

Hanna Instruments HI 83200-01 User Manual

Page 61

background image

120

121

SPECIFICATIONS

Range

0 to 150 mg/L

Resolution

5 mg/L

Accuracy

±5 mg/L ±3% of reading

Light Source

Tungsten lamp with narrow band interference filter @ 466 nm

Method

Sulfate is precipitated with barium chloride crystals. Light absorbance of the suspension

is measured.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93751-0

Indicator reagent

1 packet

REAGENT SETS

HI 93751-01 Reagents for 100 tests
HI 93751-03 Reagents for 300 tests
For other accessories see page 17.

MEASUREMENT PROCEDURE

Note: for sample preparation follow the IRRIGATION WATER (LR) procedure on page 17.

• Select the

Sulfate method using the procedure described

in the

Method Selection section (see page 12).

• Fill a cuvette with 10 mL of unreacted sample (up to

the mark) and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press ZERO key. The display will show “-0.0-” when

the meter is zeroed and ready for measurement.

• Remove the cuvette.

Sulfate

10 mL

SULFATE

Silver

• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.

• Fill a second cuvette up to the mark with 10 mL of the reacted sample.

• Insert the second cuvette into the instrument.

• Press READ to start the reading. The instrument displays

the results in mg/L of silver.

Note: a temperature above 30°C may cause turbidity. In this case, before zeroing and taking readings,
add 2-3 drops of Dispersing agent (HI 93703-51) to each cuvette and swirl until turbidity is completely

dissolved.

INTERFERENCES

Interference may be caused by:

Al

3+

above 30 mg/L

Fe

2+

above 1.5 mg/L

Ca

2+

above 1000 mg/L as CaCO

3

Fe

3+

above 10 mg/L

Cd

2+

above 20 mg/L

K

+

above 500 mg/L

Cl

-

above 8000 mg/L

Mn

2+

above 25 mg/L

Co

2+

above 1.5 mg/L

Mg

2+

above 1000 mg/L as CaCO

3

Cr

3+

above 20 mg/L

Na

+

above 5000 mg/L

Cr

6+

above 40 mg/L

Ni

2+

above 1.5 mg/L

Cu

2+

above 15 mg/L

Pb

2+

above 20 mg/L

F

-

above 20 mg/L

Zn

2+

above 30 mg/L

10 mL

# 2