• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.

• Remove the cuvette.

• Add the content of one packet of the optional reagent HI

93703-52-0 Glycine Powder. Replace the cap and shake gently

until completely dissolved.

• Add the content of one packet of HI 93757-0 Ozone Reagent.

Replace the cap and shake gently for 20 seconds.

• Replace the cuvette into the holder and close the lid.

• Press TIMER and the display will show the countdown prior to

the measurement or, alternatively, wait for 2 minutes and press

READ. When the timer ends the meter will perform the reading.

• The instrument displays a concentration value refering to chlorine interference. Subtract this value from

the reading from the Standard Measurement Procedure: this will be the concentration in mg/L of
ozone in the sample.

INTERFERENCES

Interference may be caused by: Bromine, Chlorine Dioxide, Iodine.
Alkalinity above 250 mg/L CaCO

3

will not reliably develop the full amount of color or it may rapidly fade.

To resolve this, neutralize the sample with diluted HCl.
In case of water with hardness greater than 500 mg/L CaCO

3

, shake the sample for approximately 2

minutes after adding the powder reagent.

Ozone

Ozone

• Press ZERO key. The display will show “-0.0-” the meter is zeroed and ready for measurement.

• Remove the cuvette.

• Add the content of one packet of HI 93757-0 Ozone

Reagent. Replace the cap and shake gently for 20

seconds.

• Replace the cuvette into the holder and close the lid.

• Press TIMER and the display will show the countdown

prior to the measurement or, alternatively, wait for 2
minutes and press READ. When the timer ends the

meter will perform the reading.

• The instrument displays concentration in mg/L of ozone (chlorine free samples only).

ADDITIONAL MEASUREMENT PROCEDURE

For samples containing chlorine
• Select the

Ozone method using the procedure described in

the

Method Selection section (see page 12).

• Fill the cuvette with 10 mL of unreacted sample, up to

the mark, and replace the cap.

• Place the cuvette into the holder and close the lid.