Hanna Instruments HI 931100 User Manual

Page 9

background image

17

16

• To quit the calibration and save

the first point, press the CAL
key.

For better accuracy however, it is recommended
that a two-point calibration is performed.

TWO POINT CALIBRATION

• After the first point is confirmed,

immerse the FC 300B electrode
into the second solution:
HI 7085 at 0.3 g/L NaCl or
HI 7081 at 30 g/L NaCl.

Note: The electrode should be

immersed approximately
2.5 cm (1") into the so-
lution.

• Select the second point value

by pressing the BUF key. Make
sure that the selected value cor-
responds to the one being used
for the second calibration point.

• If the reading is not close to

the selected value, the "BUF"
symbol will blink.

Otherwise, if the reading is
close to the selected value, the
meter will recognize the stan-
dard solution. Only the "

"

symbol will blink, to indicate
that the reading is stabilizing.

Once stabilized "CON" symbol
will appear on the display.

For best results, wait for a
couple of minutes and press
the CFM key.

• The two point calibration is now

completed and the meter re-
turns to the normal mode.

CAL

BUF

SINGLE POINT CALIBRATION (OFFSET)

• Dip the FC 300B sodium elec-

trode into the HI 7083 (3.00 g/L
NaCl) solution.

Note: Immerse the electrode

approx. 2.5 cm (1") into
the solution.

Press the CAL key. The "CAL" and "BUF" indi-
cators will be displayed with the "BUF" and the
"

" symbols blinking.

• As soon as the standard solu-

tion is recognized, only the
"

" symbol will continue

blinking to indicate that the elec-
trode is stabilizing.

The value of the standard solu-
tion being used will be displayed
on the primary LCD.

• As soon as the electrode is sta-

bilized, "CON" symbol lights up.

For best results, wait for a couple
of minutes before pressing the
CFM key.

HI 931100 will then display "30.0

gm".

CAL

gm

°C

NaCl

BUF

CAL

CFM

CFM