TEMPERATURE CALIBRATION

(for Technical Personnel only)

INITIAL PREPARATION

Prepare a 0°C/32°F bath with ice and water and
another one at 50°C/122°F using hot water at a
known temperature. Place insulation material around
the containers to minimize temperature changes.

Use a

Checktemp or another accurate thermometer

with a resolution of 0.1° as reference.

Note: A blinking "°C" indicates

that the temperature probe
is not connected.

PROCEDURE

• Connect the HI 7662 temperature probe to the

RCA socket on the top of the instrument.

Turn the instrument off.

To enter the temperature calibration mode, press

the CAL and ON/OFF keys simultaneously.

Immerse the temperature probe

into the 0°C/32°F bath together
with the reference thermom-
eter.

When the reading stabilizes,
press the CAL key. The
"0.0°C", "CAL" and "BUF"
symbols will be displayed. The
"

" symbol will blink for a

few seconds.

gm

°C

NaCl

0 °C

(32 °F)

°C

50 °C

(122 °F)

°C

°C

0°C

(32°F)

TEMPERATURE

The best working temperature is around 20-25°C
(68-77°F).

For greatest accuracy, it is recommended to cali-
brate the meter with standard solution at a tempera-
ture as close as possible to the temperature of the
sample (± 1°C/±2°F).

Na CONCENTRATION WITH HI 931100

It is always possible to determine the Na concentra-
tion by using the formula:

X

Na

= X

NaCl

• (23 / 58.4)

where

X

Na

= concentration of Na

and

X

NaCl

= concentration of NaCl