Hanna Instruments HI 931100 User Manual

Page 3

background image

5

4

SPECIFICATIONS HI 931100

FUNCTIONAL DESCRIPTION HI 931100

1. Battery compartment

2. BNC connector for electrode

3. Temperature probe socket

4. LCD (Liquid Crystal Display)

5. ON/OFF key, to turn the meter on or off

6. CAL key, to enter/exit calibration mode

7. CFM key, to confirm calibration

8. BUF key, to select calibration buffer value

Range

0.150 to 1.500 / 1.50 to 15.00 / 15.0 to 150.0

150 to 300 g/L NaCl

0.0 to 80.0°C

Resolution

0.001 / 0.01 / 0.1 / 1 g/L NaCl

0.1°C

Accuracy

±5% of reading (NaCl)

(@20°C)

±0.5°C

Typical EMC Deviation

±2% FS (NaCl) / ±0.5°C

Calibration

Automatic, 1 or 2 point

at 3.00 g/L (HI 7083) and

0.30 g/L (HI 7085) or 30.0 g/L (HI 7081)

Temperature Compensation

25°C (77°F)

Electrode

FC300B sodium electrode

with BNC and 1 m (3.3) cable (not included)

Temperature Probe

HI 7662 (not included)

Input Impedance

10

12

Ohm

Battery Type / Life

4 x 1.5V AA / approx. 200 hours of use

Environment

0 to 50°C (32 to 122°F);

RH max 100%

Dimensions

196 x 80 x 60 mm

(7.7 x 3.1 x 2.4")

Weight

500 g (1.1 lb.)