SINGLE POINT CALIBRATION (OFFSET)

• Dip the FC 300B sodium elec-

trode into the HI 7080 (2.3 g/L
Na) solution.

Note: Immerse the electrode

approximately 2.5 cm
(1") into the solution.

• Press the CAL key. The "CAL" and "BUF" indi-

cators will be displayed with the "BUF" and the
"

" symbols blinking.

• As soon as the standard so-

lution is recognized, only the
"

" symbol continues

blinking to indicate that the
electrode is stabilizing. The
required standard solution be-
ing value is displayed on the
primary LCD.

• As soon as the electrode sta-

bilized, the "CON" symbol ap-
pears on the display.

Press the CFM key to save
the calibration point. For best
results, wait for a couple of
minutes before confirming.

HI 931101 will display "0.230

gm".

CAL

CFM

gm

°C

BUF

CAL

Na CALIBRATION (HI 931101)

Calibration can be performed at 1 (offset only) or 2
points.

For best accuracy, a 2 point procedure is recom-
mended.

PREPARATION

• Pour small quantities of

HI7080 (2.3 g/L Na) solution
and HI 7087 (0.23 g/L Na) or
HI 7086 (23 g/L Na) into two
beakers.

For accurate readings, use for the second point
the HI 7087 solution if you are going to measure
samples with low sodium content (below 2.3 g/
L), or the HI 7086 standard if you are going to
measure samples with high sodium content (over
2.3 g/L).

• For accurate calibration use

two beakers for each solu-
tion, the first one for rinsing
the electrode, the second one
for calibration.

If possible, use plastic beakers to minimize any
EMC interferences.

• Turn the meter on by press-

ing the ON/OFF key.

• Remove the protective cap and rinse the elec-

trode with some of the calibration solution you
are going to use first.

Shift downwards the rubber
sleeve for a properly working
junction (see picture).

HI 7080

HI 7087

HI 7080

ON/OFF

HI 7080

HI 7080

RINSE

CALIBRATION