If the electrode is used rarely (once a month or
less), store it dry. Before using it again, let it
condition overnight in HI 7081 (30.0 g/L NaCl) or HI
7086
(23 g/L Na 1M)

To avoid very long response time, do not store
or rinse the electrode with deionized water. Do
not rub the electrode.

gm

°C

NaCl

TAKING pNa MEASUREMENTS (HI 931101 only)

Press the MODE key during measurement to con-
vert the value displayed by the HI 931101 to pNa.
Note that

pNa = -log a

Na

where a

Na

is the activity of ion Na.

Press the MODE key again to return to g/L mea-
surements.

TEMPERATURE MEASUREMENTS

Connect the temperature probe to
the meter and turn on.

Dip the HI 7662 temperature probe
into the sample. Allow the reading
to stabilize.

Note: The temperature is always

displayed on the secondary
LCD.

AFTER USE

Turn the meter off by pressing the
ON/OFF key again.

For faster and accurate measurements, the electrode
should be stored in a solution with a Na/NaCl concen-
tration similar to the sample to be measured.

After use, shift back upwards the
rubber sleeve (see picture).