Using the Optional Power Meters and VFL

6

Power Meter and VFL

Setting Thresholds and Correction Factors

You may apply a correction factor (CF) to measured power to compensate
for inaccuracies or drifts. You should change the CF after performing an
offset nulling.

where:

Power

corrected

= the corrected power value

Power

measured

= the measured power value

CF = the correction factor

For each favorite wavelength, the CF is set to 1.00 at the factory, but
allowed values range between 0.85 and 1.15.

Note: Some other products express the CF in dB, so the CF would be added to

measured power.

I

MPORTANT

If you revert to factory settings, you will erase all your thresholds
and correction factors. The pass/fail status feature will be disabled
automatically.

Power

corrected

Power

measured

CF

=