3 1-5v voltage output – ENMET LC-8 User Manual
Page 17
LC S
ERIES
ENMET
15
4.3.3 1-5V Voltage Output
Alarm
level 1
Alarm
Level 2
14
15
16
17
18
19
20
21
22
23
24
Channel 1
26
27
28
29
30
31
32
33
34
35
36
Channel 2
38
39
40
41
42
43
44
45
46
47
48
Channel 3
50
51
52
53
54
55
56
57
58
59
60
Channel 4
16
17
18
LC 1 15
13
25
37
49
19
20
Chart recorder
+ve
- ve
The LC Control Unit can provide a voltage output.
i.e.
zero gas
= 1V
full
scale =
5V
This output is not ideal when transmitting a signal over a large distance. The resistance of a cable attached will cause a voltage
drop to occur.
Switch ON J11, J12 and J14.
NOTE: Incorrect jumper switch configuration can cause damage to the system.
This manual is related to the following products: