ENMET LC-8 User Manual
Page 16

LC S
ERIES
ENMET
14
4.3.2
4-20mA Current Sink
The Input Module can also sink current proportional to the detected gas level.
i.e.
zero gas
= 4mA
full
scale =
20mA
The supply is derived from the external equipment.
Notes
Switch ON J10 and J13
J10
J11
J12
J13
J14
NOTE: Incorrect jumper switch configuration can cause damage to the system.