0 lel 0le w 0 lec 0 lef – ENMET EX-5100 User Manual
Page 8

ENMET
Corporation
EX-5100 C
OMBUSTIBLE
6
4.0 Operation of the EX-5100
It is best to have the
EX
-
5100
transmitters powered up and operational for 24 hours before applying calibration or test gas to
them.
When the
EX
-
5100
transmitter is first powered up, it goes through a series of momentary screens, which identify the
instrument model number, serial number and software revision. After all of the momentary screens have been displayed, the
instrument arrives at the Main Gas Display showing the gas concentration and unit of measurement.
Depending on transmitter configuration and calibration condition, the furthest right character in the display may flash a letter
indicating the instrument status. See the Section 4.1.2 below.
4.1 Start up
4.1.1 Typical Start Up
When power is supplied to the
EX
-
5100
, the S/T will display the following sequence of information:
N
OTE
:
Software revision may cause variations of display output.
Example of Display
Function
The instrument: Model
EX
-
5100
The instrument: Serial Number
The instrument: Software Revision
IF the right most character is a flashing
W
The instrument is in Warm-up mode
This should last about 1 minute
The Signal Output is held at 4mA during warm-up
For Combustible Gas
The instrument: Normal Display Mode
Measurement of target Gas
IF the right most character is a flashing
C
or
F
See Section 4.1.2
4.1.2 Alternate Start Up
Depending on
EX
-
5100
S/T configuration and calibration condition, the furthest right character in the display may flash a letter
indicating the instrument status. See the table below.
Example of Display
Function
IF the right most
character is a flashing
C
The last calibration of the instrument was invalid
The instrument must be recalibrated
IF the right most
character is a flashing
F
There is a sensor fault
To view instrument settings tap the
M
ENU
switch until 0 LEL is displayed.
EX-5100
72-1256
S/W X.X
0 LEL
0LE W
0 LEC
0 LEF