beautypg.com

3 heater voltage settings, 4 sensor replacement c, 0lel exit zero – ENMET EX-5100 User Manual

Page 15

background image

EX-5100 C

OMBUSTIBLE

ENMET Corporation

13


5.3 Heater Voltage Settings

Heater Voltages are necessary for Catalytic sensors. They are preset at the factory and should not require field adjustment. Do
not adjust these voltages unless specifically instructed to do so by

ENMET

Corporation Technical Support Staff.

C

AUTION

:

Improper adjustment of heater voltages can damage sensors voiding any warranties and also alter the operating

characteristics of the sensor in such a way that the

EX-5100

may not respond to it’s target gas.

5.4 Sensor Replacement

C

AUTION

:

Area must be declassified during sensor replacement.

Sensors should be replaced when they can no longer be calibrated. Replacement sensor part numbers are listed in Section 6.0
of this manual. If you do not know the proper part number for your sensor, be sure to have the

EX-5100

serial number

available when contacting your Distributor or

ENMET

Corporation Technical Support.

To replace a sensor, it is necessary to open the transmitter housing.

Remove the overlay and screws retaining the PC Board in the enclosure. Refer to Section 3.2, Figure 3.

Remove the sensor connector J8 and sensor

Wire in the new sensor. Refer to the wiring Table in Section 3.2, Figure 3.

After the new sensor has been installed, it is suggested to allow the sensor to stabilize for 24 hours.

A Factory calibration must be performed.

To enter the maintenance menu hold the magnet over the

M

ENU

switch for 2 to 4 seconds

The chart below indicates the maintenance menu sequence see Figure 5 for a detailed maintenance menu flow chart.

Example of Display

Function

Normal Display Mode

Measurement of target gas

Hold the magnet over the

M

ENU

switch for 2 – 4 seconds to enter the Maintenance Menu

The Power/Fault LED will flash Green – Red to indicate the

EX-5100

is in Maintenance Mode

To exit the maintenance Menu and return to the Normal Display Mode:
If intended function Tap the magnet over

S

ELECT

switch

Tap the magnet over the

M

ENU

switch to advance to the Zero procedure

For adjusting Zero:
If intended function Tap the magnet over

S

ELECT

switch

Tap the

S

ELECT

switch once with the magnet, the display will alternate between [PV: 0] and [Zero]

At this point, tap the

M

ENU

switch once with the magnet. The display should now alternate between [Zero] and

[In 300] (+ or – 30). If not then, use POT 2, to adjust to 300(+ or – 30).

N

OTE

:

The

EX-5100

will automatically reset the zero point based on a stable signal if [Span] appears before you get the 300 set

then re-enter the zero cal again, to start the clock over.

Once the zero is set you will see [Span] on the display. Tap the menu until [Z

ERO

] is displayed.

Place the magnet over the

M

ENU

switch and hold for 2-4 seconds while viewing the Zero menu.

After 2-4 seconds, an F will appear on the far right hand side of the display. The F indicates that the instrument is in Factory
mode.

Perform the calibration Zero and Span procedures as outlined in Section 5.2. Be sure that the F is present when selecting the
Zero and Span functions.

The Factory calibration sets a calibration window for future standard instrument calibrations.

Only perform a factory calibration when installing a new sensor!!

0LEL

Exit

Zero

P

OT

2 on PCB