0lel exit – ENMET EX-5100 User Manual
Page 10

ENMET
Corporation
EX-5100 C
OMBUSTIBLE
8
5.0 Maintenance of the EX-5100
C
AUTION
:
Do not open the
EX
-
5100
S/T
in a classified area.
C
AUTION
:
Do Not Attempt A Span Procedure Without Calibration Gas Applied to The Sensor; if this is done, the S/T is forced
into a calibration fault mode.
Magnetic switches control the
M
ENU
and
S
ELECT
functions. The
M
ENU
and
S
ELECT
switch locations are indicated on the
display panel, see Figure 3. The
M
ENU
switch is used to display the various menu options and make incremental changes to
numbers such as alarm points, calibrations gas, etc. The
S
ELECT
switch is used to select that option or entered digit. Most
maintenance functions are controlled by simple taps of the supplied magnet on the transmitter glass, below the
M
ENU
and
S
ELECT
boxes on the front panel.
5.1 Maintenance Menu
To enter the maintenance menu hold the magnet over the
M
ENU
switch for 2 to 4 seconds
Table 1 indicates the maintenance menu sequence see Figure 5 for a detailed maintenance menu flow chart.
Table 1: EX-5100 maintenance Menus Sequence
Example of Display
Function
Normal Display Mode
Measurement of target gas
Hold the magnet over
M
ENU
switch for 2 – 5 seconds to enter the Maintenance Menu
The Power/Fault LED will flash Green – Red to indicate the
EX-5100
is in Maintenance Mode
To exit the maintenance Menu and return to the Normal
Display Mode:
If intended function Tap the magnet over
S
ELECT
switch
Tap the magnet over the
M
ENU
switch to advance to the Zero procedure
For adjusting Zero:
If intended function Tap the magnet over
S
ELECT
switch
Tap the magnet over the
M
ENU
switch to advance to the Span procedure
For adjusting the Span:
If intended function Tap the magnet over
S
ELECT
switch
Tap the magnet over the
M
ENU
switch to advance to each Alarm set point procedures
For adjusting the Alarm 1, 2 and 3 set points:
If Intended function Tap the magnet over
S
ELECT
switch
Tap the magnet over the
M
ENU
switch to advance the mA Span set point procedure
For adjusting the mA Span set point:
If intended function Tap the magnet over
S
ELECT
switch
Taping the
M
ENU
switch without taping the
S
ELECT
switch will allow you to cycle through the menu options.
You must Tap the
S
ELECT
switch win order to change the desired operation.
N
OTE
:
If the S/T fails to respond, the magnet may have become weak and may need to be replaced.
0LEL
Exit
Zero
Span
mA Span
Alarm1
Alarm2
Alarm3