SMART LOGGER
ENMET Corporation
3
3.0 Use
3.0 Use
The operating temperature of the instrument is from 0
°
to 40
°
C.
Some gases are heavier than air, some have the same density, and some are lighter. In a closed space, hold the instrument, or
place the input of the sampling probe:
Prevent the instrument from coming in contact with any liquid, otherwise the SMART BLOCK may be damaged.
- Level with the ground for heavy gases: H
2
S, SO
2
, NO
2
, CL
2
, HCL, CLO
2
, O
3
, CH
2
OCH
2
.
- 1 meter above the ground for the detection of oxygen, CO, NO, HCN.
- High, in the case of light gases such as : H
2
, NH
3
.
3.1 Turn-On
Every 30 seconds, an audible and visual confidence signal indicates the normal operation of the instrument.
Standard turn-on:
- Press the ON/OFF switch for a moment.
The display indicates:
Then after 10 seconds:
You can read alternately:
+ The measurement (X) and the gas symbol (GAS)
+ The measurement (X) and the measurement unit, ppm or %
TEST
TEST
X - GAS
X - UNIT