3.0 Use

3.0 Use

The operating temperature of the instrument is from 0

°

to 40

°

C.

Some gases are heavier than air, some have the same density, and some are lighter. In a closed space, hold the instrument, or
place the input of the sampling probe:

Prevent the instrument from coming in contact with any liquid, otherwise the SMART BLOCK may be damaged.

- Level with the ground for heavy gases: H

2

S, SO

2

, NO

2

, CL

2

, HCL, CLO

2

, O

3

, CH

2

OCH

2

.

- 1 meter above the ground for the detection of oxygen, CO, NO, HCN.

- High, in the case of light gases such as : H

2

, NH

3

.

3.1 Turn-On

Every 30 seconds, an audible and visual confidence signal indicates the normal operation of the instrument.

Standard turn-on:

- Press the ON/OFF switch for a moment.

The display indicates:

Then after 10 seconds:

You can read alternately:

+ The measurement (X) and the gas symbol (GAS)

+ The measurement (X) and the measurement unit, ppm or %

TEST

TEST

X - GAS

X - UNIT