ENMET Corporation

SMART LOGGER

10

5.0 Features

5.0 Features

Symptoms

Cause

Solutions

Screen is blank

Used primary cell or discharged battery

Charge the primary cell or recharge the battery

No confidence beep

Out of order

Return the equipment for maintenance

Displays. Steady audible and
visual alarms no resettable

No SMART BLOCK or bad contact

Defective SMART BLOCK

Turn off the instrument, then replace the
SMART BLOCK

Change the SMART BLOCK

Displays permanently or
alternately with the measurement.
Steady or unsteady audible and
visual alarm

Primary cell or rechargeable battery life
is ending

Change the primary cell or
recharge the battery

Displays

Bad calibration
SMART BLOCK life exhausted

Turn off the instrument
Recalibrate the SMART BLOCK
Change the SMART BLOCK

Displays. Steady audible and
visual alarm

Over measurement range

Leave the area

Displays

Auto zero adjustment not performed

Reinitiate auto zero or
change SMART BLOCK

Displays

Impossible to auto zero

Stay in a safe atmosphere
Reinitiate the auto zero
Check the temperature 0 - 45

°

C

or

At turn on

Loss of time setting

Loss of memory

Change the lithium cell

Change the lithium cell

Measurements are negative

Zero not adjusted

Perform auto zero

Impossible to printout

Bad connection

Check installation

FAULT

BATTERY

USED CELL

OVERANGE

NO CAL

NOT POSS

C FAULT

C FAULT