ENMET SMARTLOGGER User Manual
Page 12

ENMET Corporation
SMART LOGGER
6
3.5 Alarms
Steady audible and visual alarm:
Display
Cause
Remedy
Low battery equipment cannot be used
anymore
Not a resettable alarm: Recharge or
charge the used battery
Measuring range exceeded
Resettable alarm: Push OPTION
switch. Leave the area.
Alternated with the O2
measurement
O2 concentration has risen or decreased
beyond alarm levels
Resettable alarm: Push OPTION
switch. Leave the area.
or
TWA or STEL alarm level have been
exceeded
Resettable alarm: Push OPTION
switch.
Alternated with gas
measurement
The instantaneous alarm level has been
exceeded
Resettable alarm: Push OPTION
switch. The alarm light stays on; the
audio alarm is OFF
Alternated with gas
measurement
Battery life is ending; 15 minutes of use
remaining
Resettable alarm: Push OPTION
switch. The alarm light stays on; the
audio alarm is OFF
As soon as an instantaneous alarm threshold is breached, the instrument displays the maximum value of the concentration
encountered during the episode, or the minimum value for oxygen deficiency. As long as the alarm is not reset with the
OPTION switch, the instrument keeps this value displayed.
OVER RANGE
ALARM
TWL AL
STEL AL
ALARM
BATTERY
BATTERY