Page 12

background image

ENMET Corporation

SMART LOGGER

6

3.5 Alarms

Steady audible and visual alarm:

Display

Cause

Remedy

Low battery equipment cannot be used
anymore

Not a resettable alarm: Recharge or
charge the used battery

Measuring range exceeded

Resettable alarm: Push OPTION
switch. Leave the area.

Alternated with the O2
measurement

O2 concentration has risen or decreased
beyond alarm levels

Resettable alarm: Push OPTION
switch. Leave the area.

or

TWA or STEL alarm level have been
exceeded

Resettable alarm: Push OPTION
switch.

Alternated with gas
measurement

The instantaneous alarm level has been
exceeded

Resettable alarm: Push OPTION
switch. The alarm light stays on; the
audio alarm is OFF

Alternated with gas
measurement

Battery life is ending; 15 minutes of use
remaining

Resettable alarm: Push OPTION
switch. The alarm light stays on; the
audio alarm is OFF

As soon as an instantaneous alarm threshold is breached, the instrument displays the maximum value of the concentration
encountered during the episode, or the minimum value for oxygen deficiency. As long as the alarm is not reset with the
OPTION switch, the instrument keeps this value displayed.

OVER RANGE

ALARM

TWL AL

STEL AL

ALARM

BATTERY

BATTERY