ENMET SMARTLOGGER User Manual
Page 15

SMART LOGGER
ENMET Corporation
9
It is possible to execute calibration by adjustment of the potentiometer located on the side of the SMART BLOCK, as
described below. This process is reserved for the users without access to the software and with time constraints which
require immediate use of the instrument. Since the potentiometer adjustment changes the calibration settings but does not
update the information in memory in the SMART BLOCK, both features and accuracy may be affected. Such a SMART
BLOCK should be returned to an authorized repair depot for recalibration with the software as soon as practical.
- execute an auto zero turn-on in a clean area
- open the SMART BLOCK cover
- place the standard gas nozzle in contact with the SMART BLOCK
- regulate the output at 1 liter per minute
- wait for stabilization of the display
– adjust the potentiometer to make the displayed value correspond with the known concentration of the gas.
4.4 Maintenance
- The instrument can become electrostatically charged, so clean only with a damp cloth.
- Recharge the battery regularly or change the primary cell out of a classified area.
- The lithium primary cell must be changed every 3 or 5 years; as follows:
For users having SMARTCOM software, the operation should be executed by authorized personnel, as follows:
- Unscrew the sensor cover screw.
- Remove the lithium primary cell which is located under the display and put a new one.
- Reprogram the date and time with SMARTCOM software.
For users who do not possess SMARTCOM software, return the instrument to your supplier or an authorized repair depot.