beautypg.com

ADLINK ND-6017 User Manual

Page 47

background image

Command Set

39

Code (Hex)

Input Range

Modules

00

±15 mV

6011/D,6018

01

±50 mV

6011/D,6018

02

±100 mV

6011/D,6018

03

±500 mV

6011/D,6018

04

±1 V

6011/D,6018

05

±2.5 V

6011/D,6018

06

±20 mA
(Required 125

Ω current conversion resistor.)

6011/D,6018

08

±10 V

6012/D,6017,
6014D

09

±5 V

6012/D,6017,
6014D

0A

±1 V

6012/D,6017,
6014D

0B

±500 mV

6012/D,6017,
6014D

0C

±150 mV

6012/D,6017,
6014D

0D

±20 mA
(Required 125

Ω current conversion resistor.)

6012/D,6017,
6014D

0E

Type J Thermocouple 0

° to 760°C

6011/D,6018

0F

Type K Thermocouple 0

° to 1370°C

6011/D,6018

10

Type T Thermocouple -100

° to 400°C

6011/D,6018

11

Type E Thermocouple 0

° to 1000°C

6011/D,6018

12

Type R Thermocouple 500

° to 1750°C

6011/D,6018

13

Type S Thermocouple 500

° to 1750°C

6011/D,6018

14

Type B Thermocouple 500

° to 1800°C

6011/D,6018

15

Type N Thermocouple -270

° to 1300°C

6011/D,6018

16

Type C Thermocouple 0

° to 2320°C

6011/D,6018

20

Pt-100, -100

°C to +100°C, α=0.00385

6013

21

Pt-100, 0

°C to +100°C, α=0.00385

6013

22

Pt-100, 0

°C to +200°C, α=0.00385

6013

23

Pt-100, 0

°C to +600°C, α=0.00385

6013

24

Pt-100, -100

°C to +100°C, α=0.003916

6013

25

Pt-100, 0

°C to +100°C, α=0.003916

6013

26

Pt-100, 0

°C to +200°C, α=0.003916

6013

27

Pt-100, 0

°C to +600°C, α=0.003916

6013

28

Ni-100, 0

°C to +100°C

6013

29

Ni-120, 0

°C to +100°C

6013

2A 0~60

Ohms

6013

Table 3-1 AD Input Range Setting