5 acquisition with re-triggering, Acquisition with re-triggering, Figure 3-9 – ADLINK PCIe-9814 User Manual
Page 31: Middle trigger mode acquisition, Figure 3-10, Re-trigger mode acquisition
Operations
21
PCIe-9814
Figure 3-9: Middle Trigger Mode Acquisition
3.4.5
Acquisition with Re-Triggering
A digitizer acquires a trace of N samples/channel for a single
acquisition. Re-Trigger mode can also be set to automatically
acquire R traces, containing N*R samples/channel of data, without
additional software intervention.
The Re-Trigger setting can be used for Post-Trigger and Delayed-
Trigger modes, with different limitations on the spacing between
trigger events in each mode. Trigger events arriving too close to
the previous instance will be ignored by the digitizer.
X
In Post-Trigger mode, the minimum spacing between trigger
events is N+1
X
In Delayed-Trigger mode, the minimum spacing between
trigger events is (N+D)+1, where D is the number of the
delayed setting
Figure 3-10: Re-Trigger Mode Acquisition
Time
Operation start
Acquisition start
Trigger
Data
Acquisition stop
Data transfer to system begins
N samples
M samples
Trigger event occurs