Vacuum management with vms module a – VACUUBRAND CVC 2II User Manual
Page 16
page 16 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 997835 / 15/12/2009
Attention: The basic setups Management, Management
Plus or VACUU•LAN are selectable only if a
VMS Module A and an isolation valve if necessary are connected!
Notes on choosing the basic setup
The controller CVC 2
II
can be adapted to the specific application by choosing the
appropriate basic setup Standard, Management, Management
Plus or VACUU•LAN
and choosing an operation mode, two-point control or continuous pumping: The con-
nected components of the chemistry vacuum system have to be preset once only.
The vacuum controller CVC 2
II
offers four basic setups depending on
-
the components of the VACUUBRAND chemistry vacuum system which are con-
nected to the system.
-
the specific user and/or process requirements.
The controller CVC 2
II
identifies automatically if an isolation valve is connected. If it is
missing, the basic setups “Standard“ and “Management
Plus“ are not selectable.
Warning triangle and symbol isolation valve are flashing:
➨ Isolation valve is missing, is however indispensable for the
selected basic setup.
☞ Connect isolation valve or choose appropriate basic setup.