Important notes regarding the use of gas ballast, Notice – VACUUBRAND MV 10C NT + EK User Manual
Page 39

page 39 of 92
In case of overload, the motor is shut down by
a
self-hold thermal cutout in the winding.
Note: Only manual reset is possible. Switch
off the pump and disconnect from the power
source. Identify and eliminate the cause of fail-
ure. Allow the pump to cool down sufficiently
before restart.
°C / °F
I ➨ O
!
3.
2.
1.
IN
✘
Avoid overheating (e.g., due to hot process
gases). Maximum permissible temperature
range: see ”Gas inlet temperatures”, pg. 26.
Make sure the fan’s air supply is adequate. Do
not place the pump on soft surfaces (e.g., rub-
ber foam) during operation. This may reduce
the air circulation in the pump housing and af-
fect the fan’s cooling effect. Check fan regu-
larly for dust/dirt. Clean fan guard grill if neces-
sary to avoid a reduction of ventilation.
air
NOTICE
NOTICE
NOTICE
A warm up period (approximately 15 min.) is
required to ensure that the rated ultimate vac-
uum and pumping speed are attained.
O
➨ I
m
3
/h
Torr/mbar
✔
✔
~15min
Important notes regarding the use of gas ballast
Gas ballast is a continuous purge to keep the pump’s in-
terior as clean as possible and to reduce the possibility of
condensation inside the pump.