Electrical connection, During operation, Notice – VACUUBRAND MV 10C NT + EK User Manual
Page 37: Electrical connection during operation
page 37 of 92
Permissible range of coolant temperature at the exhaust
waste vapor condenser: 5°F to 68°F (-15°C to +20°C)
Check hose connections prior to starting operation of the
cooling system.
Check coolant hoses regularly during operation.
NOTICE
Plug in the power cord.
• Check the power source and the pump’s
rating plate to be sure that the power source
and the equipment match in voltage, phase,
and frequency.
Switch the pump on.
O
➨ I
Electrical connection
During operation
➨
Vent and dispose of
potentially dangerous gases or
vapors at the outlet of the pump appropriately.
+
Due to the high compression ratio, the pump might
generate overpressure at the outlet. Check pressure
compatibility with system components (e.g., exhaust
tubing or exhaust valve) at the outlet. Ensure that the
pump outlet is neither blocked nor restricted.
+
Maximum ambient temperature: 104 °F
(40 °C)
Check the maximum temperatures, if install-
ing the pump in a cabinet or a housing.
Make sure ventilation is adequate, especial-
ly if the ambient temperature is elevated.
max.
104°F
(40°C)