Technical data
General technical data valid for all pumps
Maximum permissible inlet pressure
(absolute)
psi
(bar)
16
(1.1)
Maximum permissible outlet pressure
(absolute)
psi
(bar)
16
(1.1)
Maximum pressure difference between
inlet and outlet
psi
(bar)
16
(1.1)
Maximum permissible pressure (abso-
lute) at gas ballast valve
psi
(bar)
17.5
(1.2)
Permissible ambient temperature
storage / operation
°F
(°C)
14 to 140 / 50 to 104
(-10 to +60 / +10 to +40)
Permissible relative atmospheric moisture
during operation (no condensation)
%
30 to 85
Maximum permissible installation altitude
above mean sea level
ft
(m)
6500
(2000)
Rated motor power
hp
(kW)
0.59
(0.44)
No-load speed 50/60 Hz
rpm
1500 / 1800
Maximum permissible range
of supply voltage (±10%)
Attention: Observe specifications of rating plate!
100-115 V~ 50/60 Hz,
120 V~ 60 Hz
230 V~ 50/60 Hz
Maximum rated current at:
100-115 V~ 50/60 Hz, 120 V~ 60 Hz
230 V~ 50/60 Hz
A
A
8.0
3.0
Maximum start-up current
100-115 V~ 50/60 Hz, 120 V~ 60 Hz
230 V~ 50/60 Hz
A
A
15.5
6.1
Device fuse
2 slow blow fuses 8A
Motor protection
thermal cutout, manual reset
Degree of protection IEC 529
IP 40
Inlet
small flange KF DN 25
Outlet
+ EK
hose nozzle for tubing I.D. 5/8”
(hose nozzle DN 15 mm)
hose nozzle for tubing I.D. 3/8”
(hose nozzle DN 10 mm)