Decontamination by autoclaving
121°C
2 bar absolute
Autoclaving
The collection bottle with bottle head and screw cap, the quick
coupling, the filter and the VHC
pro
with adapters and aspirati-
on hose are designated for steam sterilization at 121°C and
2 bar absolute (1 bar overpressure). Time of exposure accor-
ding to DIN 58946 t
e
= 20 minutes.
Ensure not to exceed a temperature of 121°C.
Comply with instructions for use of the manufacturer of the
autoclave.
It is the user´s responsibility to ensure effective autoclaving.
Prior to autoclaving loosen or remove the bottle head
from the bottle.
Midi
sart® 2000
0.20
µm PTFE
IN
sartorius stedim
Prior to autoclaving set the VHC
pro
to “continuous aspira-
tion / autoclaving”.
The applicability of the decontamination method has to be ensured by the user.
Prior to decontamination clean surfaces with a damp cloth if necessary. Use water or
a mild detergent as cleaning agent.