VACUUBRAND VHC Pro User Manual
Page 23
page 23 of 27
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 25/11/2013 // 999281
➨
Screw in hose nozzle or hose nozzle with coupling using
an open-ended wrench.
➨
Screw on cover, paying attention to washers.
Maximum torque: 2.2 ft
.
lb
f
(3 Nm).
➨
Only BVC professional / G: In case the cover plate is shift-
Only BVC professional / G: In case the cover plate is shift-
ed, ensure that the ribbon cable is not squeezed during
reassembly.
➨
Position bottle in the support.
➨
Position tubing to pump inlet and filter.
➨
Screw elbow fitting with silencer to the outlet of the pump
(max. 5 turns).
➨
Screw elbow fitting with silencer to the outlet of the pump
(max. 5 turns).