beautypg.com

Decontamination of the pump due to gassing, Notice – VACUUBRAND VHC Pro User Manual

Page 14

background image

page 14 of 27

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. 25/11/2013 // 999281

Connect gas pressure bottle to the pump inlet.

+

Keep the valve of the gas pressure bottle closed.

Remove hose at the pump inlet.

Switch off the BVC.

Position the BVC in a hood or install an exhaust pipe at the

outlet of the pump to ensure proper disposal of outgoing

gas.

The applicability of the decontamination method and the selection of the ap-

propriate gas (e. g., formaldehyde, ethylene oxide, ammonia, .....) have to be

ensured by the user.

Decontamination of the pump due to gassing

NOTICE

This manual is related to the following products: