Page 36: Hard cutout, Ct1, ct2, and ct3, Ce1, ce2, and ce3

CAL
Calibration parameters menu
Press “SET” five times to enter the menu. The calibration parameters menu
contains the parameters Hard Cutout, CT1, CE1, CT2, CE2, CT3, and CE3.
8.12.1
Hard Cutout
This parameter is the temperature above which the unit shuts down automati-
cally. The parameter is set at the factory to approximately 1260°C and cannot
be changed by the user.
8.12.2
CT1, CT2, and CT3
The calibration parameters CT1, CT2, and CT3 are the calibration
temperatures.
8.12.3
CE1, CE2, and CE3
The calibration parameters CE1, CE2 and CE3 are the calibration errors corre-
sponding to the calibration temperatures.
35
8 Controller Operation
Calibration Parameters
See also other documents in the category ThermoWorks Measuring instruments:
- LIMITED EDITION THERMAPEN (4 pages)
- LIMITED EDITION THERMAPEN Calibration Procedure (2 pages)
- SPLASH-PROOF SUPER-FAST THERMAPEN (4 pages)
- THERMADATA (RF) WIRELESS TEMPERATURE & DATA LOGGERS Studio (1 page)
- THERMADATA TEMPERATURE & HUMIDITY LOGGERS – SERIES II (1 page)
- TRIX-8 LOGTAG TRANSIT RECORDERS Quick Start Guide (2 pages)
- HACCP MOBILE FOR USE WITH BLUETHERM BLUETOOTH TEMPERATURE PROBE (1 page)
- 222-213 SPLASH-PROOF REFERENCE THERMAPEN (2 pages)
- 231-214 AIR THERMAPEN (2 pages)
- 231-279 PRO-SURFACE THERMAPEN (2 pages)
- RTR-61 WIRELESS HACCP RECORDING THERMOMETER (2 pages)
- 9882 PRINTING DATALOGGER THERMOMETER (54 pages)
- P750 PRINTING DATALOGGER THERMOMETER (28 pages)
- P795 PRINTING DATALOGGER THERMOMETER (9 pages)
- 222-550 PRECISION PLUS THERMOMETER (1 page)
- 232-042 THERMACHECK PLUS PRECISION THERMOMETER (4 pages)
- 226-042 THERMACHECK (2 pages)
- 292-701 SAF-T-LOG PAPERLESS HACCP THERMOMETER Quick Start Guide (4 pages)
- 292-701 SAF-T-LOG PAPERLESS HACCP THERMOMETER Operating Instructions (18 pages)
- 221-059 CATERTEMP PLUS (2 pages)
- 231-022 THERMA DIFFERENTIAL THERMOCOUPLE METER (2 pages)
- 232-101 THERMA WATERPROOF THERMOMETER (4 pages)
- 222-910 TEMPTEST 2 (2 pages)
- 221-051 THERMA K-PLUS (2 pages)
- 221-048 FOOD CHECK THERMOCOUPLE SYSTEM (2 pages)
- 261-550 THERMAMITE THERMOCOUPLE METER WITH FIXED PENETRATION PROBE (4 pages)
- TW8060 TWO-CHANNEL THERMOCOUPLE WITH ALARM (16 pages)
- CF-IR CLOSE FOCUS INFRARED THERMOMETER (2 pages)
- R-PRO PROFESSIONAL INFRARED THERMOMETER (2 pages)
- R-IND INDUSTRIAL INFRARED THERMOMETER WITH CIRCLE LASER Operating Instructions (2 pages)
- R-IND INDUSTRIAL INFRARED THERMOMETER WITH CIRCLE LASER Targeting Guide (1 page)
- IRK-2 INFRARED THERMOMETER Operating Instructions (4 pages)
- IRT INFRARED + TYPE T INPUT THERMOMETER (2 pages)
- 814-065 COMBO THERMOMETER INFRARED WITH FOLD OUT PROBE (3 pages)
- IRFS INFRARED FOOD SAFETY THERMOMETER (2 pages)
- IR-GUN-S INDUSTRIAL INFRARED THERMOMETER (2 pages)
- TW2 POCKET INFRARED THERMOMETER (1 page)
- IR-POCKET POCKET INFRARED THERMOMETER (1 page)
- 292-601 BLUETHERM DUO WITH BLUETHERM PRO FOR iOS Operating Instructions (2 pages)
- 292-601 BLUETHERM DUO WITH BLUETHERM PRO FOR iOS Quick Start Guide (1 page)
- 810-960 ECOTEMP ALARM THERMOMETER (1 page)
- PHARMALARM 2 MAX/MIN RECORDING THERMOMETER W/CAL CERTIFICATE (1 page)
- PHARMALARM 1 DIGITAL ALARM THERMOMETER W/ CAL CERTIFICATE (1 page)
- RT8100 FRIDGE/FREEZER ALARM THERMOMETER (1 page)
- RT8100MAT MIN/MAX RECORDING THERMOMETER WITH ALARMS (1 page)