beautypg.com

Service information, Section iii – service information – Sonics VCF1500HV User Manual

Page 14

background image









SECTION III – SERVICE INFORMATION




Your Ultrasonic Processor was designed to provide you with years of safe and
dependable service. Nevertheless, because of component failure or improper usage, the
possibility does exist that it might not perform, as it should, shut down due to an overload
condition or that it will stop working all together. The most probable causes for
malfunction are listed below and should be investigated.

The unit was plugged into an electrical outlet that provides a different voltage from
that required. See Electrical Requirements.

The probe is not secured properly to the converter.

A fuse(s) has failed. If a fuse(s) has failed, proceed as follows:

1.

Ensure that the

POWER

switch is set to OFF.

2.

Replace the fuse(s).

3.

Drain the High Volume Flow Cell completely.

4.

Set the AMPLITUDE control to 50, the

POWER

switch back to ON, and depress

the

TEST

switch. The power monitor should read below 20%. If the reading

exceeds 20%, set the

POWER

switch to OFF, and disconnect the probe from the

converter.

5.

Set the

POWER

switch back to ON and depress the

TEST

switch. If the power

monitor reads below 20%, the probe has failed or is out of tune due to excessive
erosion, and should be replaced, if the power monitor reads above 20%, either the
converter or power supply has failed. The Ultrasonic Processor together with the
High Volume Flow Cell should be returned for repair.

14

Go To Top Of Document