CAUTION
Never allow liquid to spill into the converter.
1.
With the liquid flowing through the High Volume Flow Cell, adjust the cell pressure
and Amplitude control to achieve 60% reading on the power monitor. If the liquid is
discharged into an open container or a low-pressure line, the pressure can be
increased by reducing the flow rate with a valve on the discharge side. Unless
absolutely necessary do not operate the power supply continuously with the
Amplitude control set above 90%.
2.
Set the
TIMER
as required
3.
Depress the
START
button.
4.
Using the AMPLITUDE control, increase or decrease the amplitude as required.
5.
If the converter runs hot, air-cool with DRY compressed air.
6.
Process the material at different flow rates, while maintaining constant pressure.
7.
Evaluate the sample.
8.
If the sample is over-processed or the temperature reaches an unacceptable level,
reduce the amplitude.
NOTE
When working with high viscosity material, it is recommended that (if possible) the
viscosity be reduced as much as possible prior to processing.
11