beautypg.com

Slant/Fin SC-3 User Manual

Page 4

background image

5

SC-3 Controller

WIRING AND TESTING THE SENSOR

WARNING: Do not run sensor wires close to and parallel to telephone or power cables. If the sensor wires are located in an area with

strong sources of electromagnetic interference, shielded cable or twisted pair should be used or the wires can be run in a grounded metal

conduit. If using shielded cable, the shield wire should be connected to the Com Sen terminal on the control and not to earth ground.

• It is necessary to connect 18 AWG wire to the two sensor wires. Wire nuts can be used to hold the wires together.

• Follow the sensor testing instructions that follow and then connect the wires to the control.

SENSOR TESTING INSTRUCTIONS

A good quality test meter capable of measuring up to 5,000 k

Ω (1 kΩ = 1000 Ω) is required to measure the sensor resistance. In addition

to this, the actual temperature must be measured with either a good quality digital thermometer, or if a thermometer is not available,

a second sensor can be placed alongside the one to be tested and the readings compared.

First measure the temperature using the thermometer and then measure the resistance of the sensor at the control. The wires from the

sensor must not be connected to the control while the test is performed. Using the chart below, estimate the temperature measured by

the sensor. The sensor and thermometer readings should be close. If the test meter reads a very high resistance, there may be a broken

wire, a poor wiring connection or a defective sensor. If the resistance is very low, the wiring may be shorted, there may be moisture in

the sensor or the sensor may be defective. To test for a defective sensor, measure the resistance directly at the sensor location.

Do not apply voltage to a sensor at any time as damage to the sensor may result.

Temperature

Resistance

Temperature

Resistance

Temperature

Resistance

Temperature

Resistance

˚F

˚C

˚F

˚C

˚F

˚C

˚F

˚C

-50

-46

490,813

20

-7

46,218

90

32

7,334

160

71

1,689

-45

-43

405,710

25

-4

39,913

95

35

6,532

165

74

1,538

-40

-40

336,606

30

-1

34,558

100

38

5,828

170

77

1,403

-35

-37

280,279

35

2

29,996

105

41

5,210

175

79

1,281

-30

-34

234,196

40

4

26,099

110

43

4,665

180

82

1,172

-25

-32

196,358

45

7

22,763

115

46

4,184

185

85

1,073

-20

-29

165,180

50

10

19,900

120

49

3,760

190

88

983

-15

-26

139,402

55

13

17,436

125

52

3,383

195

91

903

-10

-23

118,018

60

16

15,311

130

54

3,050

200

93

829

-5

-21

100,221

65

18

13,474

135

57

2,754

205

96

763

0

-18

85,362

70

21

11,883

140

60

2,490

210

99

703

5

-15

72,918

75

24

10,501

145

63

2,255

215

102

648

10

-12

62,465

80

27

9,299

150

66

2,045

220

104

598

15

-9

53,658

85

29

8,250

155

68

1,857

225

107

553

The Universal Sensor can be

strapped directly to the pipe using

the cable tie provided. Insulation

should be placed around the sensor

to reduce the effect of air currents on

the sensor measurement. The flat

portion of the sensor should be mounted

against the pipe.

The Universal Sensor should be placed

downstream of a pump or after an elbow

or similar fitting. This is especially impor-

tant if large diameter pipes are used as

the thermal stratification within the pipe

can result in erroneous sensor readings.

Proper sensor location requires that the

fluid is thoroughly mixed within the pipe

before it reaches the sensor.