*RCL_
This is the IEEE488.2 common command. It takes the single numeric argument [1-4], and restores the
instrument configuration to that configuration stored in the corresponding save area. It also places the IEEE
interface into the OCIS and OQIS states.

COUNT_
Sets the number of pulses making up each average in average energy mode.

COUNT?
Returns the number of pulses making up each average in average energy mode.

CALIB_
Sets the calibration constant (Volts/Joule) for a pyro detector, (Volt/Watt) for an Ultra Series detector or the
time constant for a calorimeter. Note: This command is context sensitive; the detector type must be set by the
SENS command before this command is executed.

CALIB?
Returns the constant for a pyro detector, or the time constant for a calorimeter. Note: This command is context
sensitive; the returned value depends on the detector type set by the SENS command.

ATTEN_
Sets an attenuation factor for the current detector.

ATTEN?
Returns the attenuation factor for the current detector.

ZERO
This command zeros the power baseline for a calorimeter.

PSWD_
Enables password protection commands if the password is entered correctly.

PSPEED HF|BL
Sets the electronic response speed according to the repetition rate capability of the pyroelectric detector type - a
"black" coated absorbing crystal, or uncoated "high frequency" absorbing crystal.

PSPEED?
Returns the pyro speed, possible responses are BL and HF.

CSPEED_
Sets the watts mode display response speed for calorimeters and Ultra Series detectors.

CSPEED?
Returns the watts mode display response speed for calorimeters and Ultra Series detectors.

CDELAY_
Sets the "Calorimeter Delay" function of the meter.

CDELAY?
Returns the "Calorimeter Delay" function of the meter.